Home > Contents

Article


Biomed Imaging Interv J 2010; 6(3):e31
doi: 10.2349/biij.6.3.e31
© 2010 Biomedical Imaging and Intervention Journal


PDF version Technical Report

Optimal slice thickness for cone-beam CT with on-board imager

Seet KYT, PhD, Barghi A, Yartsev S*, PhD, MCCPM, Van Dyk J, FCCPM, DABMP, FAAPM, FInstP
London Regional Cancer Program, London Health Sciences Centre, London, Ontario, Canada
* Corresponding author. Present address: Department of Physics and Engineering, London Regional Cancer Program, London Health Sciences Centre, 790 Commissioners Road East, London, Ontario, Canada N6A 4L6. Tel.: +1-5196858605; Fax: +1-5196858658; E-mail: slav.yartsev@lhsc.on.ca (Slav Yartsev).
Received 21 January 2010; received in revised form 18 May 2010; accepted 19 May 2010

ABSTRACT

Purpose: To find the optimal slice thickness (Δτ) setting for patient registration with kilovoltage cone-beam CT (kVCBCT) on the Varian On Board Imager (OBI) system by investigating the relationship of slice thickness to automatic registration accuracy and contrast-to-noise ratio.

Materials and method: Automatic registration was performed on kVCBCT studies of the head and pelvis of a RANDO anthropomorphic phantom. Images were reconstructed with 1.0 ≤ Δτ (mm) ≤ 5.0 at 1.0 mm increments. The phantoms were offset by a known amount, and the suggested shifts were compared to the known shifts by calculating the residual error. A uniform cylindrical phantom with cylindrical inserts of various known CT numbers was scanned with kVCBCT at 1.0 ≤ Δτ (mm) ≤ 5.0 at increments of 0.5 mm. The contrast-to-noise ratios for the inserts were measured at each Δτ.

Results: For the planning CT slice thickness used in this study, there was no significant difference in residual error below a threshold equal to the planning CT slice thickness. For Δτ > 3.0 mm, residual error increased for both the head and pelvis phantom studies. The contrast-to-noise ratio is proportional to slice thickness until Δτ = 2.5 mm. Beyond this point, the contrast-to-noise ratio was not affected by Δτ.

Conclusion: Automatic registration accuracy is greatest when 1.0 ≤ Δτ (mm) ≤ 3.0 is used. Contrast-to-noise ratio is optimal for the 2.5 ≤ Δτ (mm) ≤ 5.0 range. Therefore 2.5 ≤ Δτ (mm) ≤ 3.0 is recommended for kVCBCT patient registration where the planning CT is 3.0 mm. © 2010 Biomedical Imaging and Intervention Journal. All rights reserved.

Keywords: Cone beam CT, slice thickness, automatic registration, contrast-to-noise ratio, on-board imager

INTRODUCTION

Image-guided radiation therapy (IGRT) is carried out to improve treatment accuracy in patients. The Varian on-board imaging (OBI) system is an IGRT modality that permits kilovoltage cone-beam CT on the treatment machine. kVCBCT obtains 3D volumetric studies of patients in the treatment room immediately before radiation delivery. The main advantage of cone-beam CT is the minimisation of patient movement between imaging and treatment, such that dose gradients conform to planning target volumes. These images, although prone to artefacts such as streaking and cupping effects, can still be used to ensure accurate and consistent patient positioning prior to treatment. kVCBCT is an ideal tool for fast patient registration, only requiring one 360° rotation of the gantry (t = 60 s) for acquisition. Both the kVCBCT and treatment beam (MV) are commissioned to share the same isocentre. The OBI software performs an automatic registration of cone-beam CT studies to match planning CT and the patient can be shifted accordingly with reference to the isocentre. Previous investigations on the usage of megavoltage CT (MVCT) studies for daily set-up on helical tomography have found the automatic registration process to be an adequate procedure [1, 2]. In volumetric CT studies, the axis of gantry rotation is referred to as the longitudinal axis. The resolution along this axis is referred to as the ‘slice thickness’ (Δτ) of the CT study. In planning CT, Δτ has a significant effect on how much radiation is delivered to the patient during the imaging session [3]. Smaller Δτ can therefore only be achieved at the price of increased imaging dose in planning CT. To increase resolution, interpolation of acquired planning CT data is possible, but it causes aliasing artefacts [4]. Therefore, the longitudinal direction has the lowest resolution in planning CT studies. Unlike planning CT, kVCBCT always acquires all its information in a single gantry rotation regardless of slice thickness. Selecting a different kVCBCT slice thickness only affects how the information is partitioned from the flat-panel detector during reconstruction, and has no influence on the method of image acquisition or amount of radiation delivered by the x-ray generator. For reconstruction of the kVCBCT image using the Varian OBI system, Δτ ranging from 1 to 5 mm can be selected at increments of 0.5 mm. This study focuses on automatic registrations using planning CT at 3 mm slice spacing, to evaluate the consequences, if any, of increasing Δτ greater than that of the reference set, as well as the potential advantages of using Δτ smaller than the resolution of the reference set. A motivation for this investigation comes from megavoltage CT investigations on automatic registration accuracy, which have found that decreasing slice thickness past a certain value offers no advantage since registration accuracy becomes very similar [1]. Sykes et al. investigated automatic registration of kVCBCT with planning CT of various beam energy and slice thickness settings for the Synergy system (Elekta, UK) [5]. At the London Regional Cancer Program (LRCP), a standard 3 mm slice thickness is used clinically for planning CT imaging. This study aims to explore the optimal Δτ value for kVCBCT image reconstruction to achieve optimal image registration with 3 mm planning CT studies and image quality (defined by contrast-to-noise ratio).

Cone beam images generally produce higher noise (standard deviation in HU) in kVCBCT studies compared to planning CT studies when imaged with a similar set of scan parameters. Studies have shown that noise levels seen in kVCBCT images do not cause a significant loss in automated registration accuracy [6]. However, for the manual registrations, Δτ choice is important because the apparent “visibility” can be improved with an optimal setting, making image noise an important factor to investigate when considering Δτ optimisation. Resolution and noise trade-offs have been observed in previous investigations with fan-beam CT [7, 8]. Lowering the resolution by increasing Δτ may reduce noise in kVCBCT images in cases where longitudinal variation in CT numbers is minimal. If automatic registration accuracy is not affected by slice thickness, as is the case in MVCT investigations, reducing noise with kVCBCT would be possible with no loss of automatic registration accuracy.

MATERIALS AND METHODS

Registration accuracy vs. slice thickness

The OBI system consists of a kV x-ray source and flat panel detector mounted orthogonal to the treatment beam axis. For this study, registration accuracy was investigated by performing automatic registrations to the head and pelvis of a RANDO anthropomorphic male phantom, manufactured by The Phantom Laboratory (Salem, NY). The RANDO phantom is a 73.5 kg male human mould of synthetic soft tissue-equivalent material with real human skeleton. Planning CT studies of the phantom were acquired with a Philips PQ5000 CT simulator with 120 kV, 85 mA, 3 mm slice spacing, 25 cm FOV and 512 × 512 slice resolution (typical clinical setting) and transferred to the Eclipse planning system via the RT DICOM protocol. This investigation focused on 3 mm slice spacing for planning CT as the reference set following the authors’ institution practice, which is the median value of Δτ for OBI reconstruction. The head and pelvis of the phantom were aligned to the isocentre of the OBI system using the room’s lasers and metal ball bearings that were attached to the phantom during planning CT. The phantoms were then offset by a known amount in the lateral (Δx), superior-inferior (Δy), and anterior-posterior (Δz) directions. In this offset position, a kVCBCT study was acquired at 125 kVp, 80 mA, and 25 ms. Fields of view of 25 × 25 cm2 and 50 × 50 cm2 were used for the head and pelvis of the phantom, respectively. Slices were reconstructed at 512 × 512 pixels. A bowtie filter, which is an aluminium attachment that covers the x-ray source, was used to reduce artefacts, as recommended by the manufacturer.

With the Varian OBI software, kVCBCT and planning CT studies can be viewed overlapping one another. The user has the option of manual or automatic registration. The automatic registration algorithm uses a similarity measure cost function to find the global maximum to match pixel intensities [9]. Correctional shifts are displayed with 1 mm accuracy, which is the same as the resolution of the couch position. For five values of slice thickness Δτ ranging from 1 to 5 mm, automatic matches were performed and the suggested shifts were recorded and compared to the known shifts for both the head and pelvis of the phantom. Residual error was calculated using:

Equation 1

where R is the residual error, xt, yt, and zt are the suggested shifts given by the automatic match, and Δx, Δy, and Δz are the known shifts. To account for set-up errors, phantoms were returned to an offset position and the automatic registration process repeated four times. The residual error was then averaged.

Image noise vs. slice thickness

A CatPhan phantom, a cylindrical phantom of radius 10 cm containing cylindrical inserts of radius 0.6 cm and of varying densities, manufactured by The Phantom Laboratory (Salem NY), was scanned with kVCBCT using a bowtie filter at 125 kVp, 80 mA, and 25 ms, and reconstructed at 5-slice thickness in the range of 1 ≤ Δτ (mm) ≤ 5. Inserts with known CT numbers (−1000, −200, −100, 340, and 990 HU) were compared to the CatPhan’s uniform background of 100 HU. Using ImageJ [10], an image-processing program developed by The National Institutes of Health, the CT numbers of the inserts (on the central slice) and their surrounding background were measured. For each insert, the average CT number and standard deviation of a circular area of radius 0.6 cm was determined. A ring of outer radius 1.0 cm and inner radius 0.8 cm was used around each insert to characterise the background. Contrast-to-noise ratios were calculated using relation:

Equation 2

where CNR is the contrast-to-noise ratio, HUins is the average CT number of the insert, HUsurr is the average CT number of the surroundings, σins is the standard deviation of the CT number of the insert, and σsurr is the standard deviation of the CT number of the surroundings.

RESULTS AND DISCUSSION

The residual errors of the automatic registrations calculated using eq. (1) from the measurements on the head and pelvis of the RANDO phantom for varying Δτ are presented in Figures 1(a) and 1(b), respectively. For Δτ ≤ 3 mm, there is no significant difference in residual error. For Δτ > 3 mm, residual error increases. The increase in residual error is greatest when the initial translation is in the longitudinal direction. This is the direction in which changing Δτ changes the resolution. Δτ = 3.0 mm is the limit at which Δτ can be increased without affecting the accuracy of the automatic registration. This is because the planning CT was taken at a slice spacing of 3.0 mm, which suggests that there is no advantage for automatic registration to use a resolution lower than that of the reference set. This may be a consequence of the registration algorithm, which is limited by the finite axial resolution of the planning CT. Reconstruction occurs before automatic registration, meaning that CBCT data is reconstructed into finite axial slices before the automatic registration algorithm is used, but the algorithm can interpolate the CBCT data to achieve maximum overlap. However, the planning CT slices are fixed during this process, so registration accuracy will not improve past a certain point by changing the resolution of the CBCT data alone. These results show that the automatic registration procedure is optimal when Δτ ≤ 3 mm, with no advantage of one slice thickness over another within this range.

Residual errors vs. slice thickness of the RANDO (a) head phantom and (b) pelvis phantom registered at initial translations in the x, y, and z directions. Figure 1 Residual errors vs. slice thickness of the RANDO (a) head phantom and (b) pelvis phantom registered at initial translations in the x, y, and z directions.

The contrast-to-noise ratio was determined by the measurement of insert of various CT numbers. Figure 2 shows the dependence of contrast-to-noise ratio on slice thickness for five different inserts within a CatPhan phantom. CNR increases steadily within the range 1.0 ≤ Δτ (mm) ≤ 2.5, then saturates for greater values. This is due to the low resolution of the kVCBCT flat panel detector with rectangular pixels of 0.2 × 0.4 mm2. A reduction of pixels in one direction also affects the smoothing of noise.

Contrast-to-noise ratio vs. slice thickness for inserts of air, polymethylpentene, low-density polyethylene, Delrin, and Teflon within a CatPhan phantom. Figure 2 Contrast-to-noise ratio vs. slice thickness for inserts of air, polymethylpentene, low-density polyethylene, Delrin, and Teflon within a CatPhan phantom.

Figure 3 shows kVCBCT slices of the CatPhan phantom with low contrast inserts reconstructed at 1.0 mm and 2.5 mm slice thicknesses. This demonstrates the significance of the CNR decrease seen at low Δτ with kVCBCT, where noise levels in the 1.0 mm reconstructed slice are high enough that the visibility of some smaller inserts becomes less. Reconstruction of the CatPhan at Δτ ≥ 2.5 mm therefore improved visibility of low contrast structures.

kVCBCT slices of CatPhan phantom with low contrast inserts reconstructed at (a) 1.0 and (b) 2.5 mm slice thickness. Window and level are the same for both images. Figure 3 kVCBCT slices of CatPhan phantom with low contrast inserts reconstructed at (a) 1.0 and (b) 2.5 mm slice thickness. Window and level are the same for both images.

Inserts were also located within the central portion of the phantom where the background HU remained constant (within ±5 HU). Depending on the field of view and object size, reconstructed images can exhibit streaking and cupping artefacts, resulting in CT number inaccuracy [11]. In this study, measurements were not affected by these artefacts due to imaging parameters and phantom size.

In this study, pCT images of the head and pelvis of the RANDO phantom were captured at 3.0 mm slice thickness as this was the clinical standard for slice thickness. Decreasing this slice thickness is likely to produce registrations of greater accuracy. However, results in this study demonstrate the best outcome for least accuracy in slice thickness. One of the new features of the recent OBI version1.4 is the automatic registration with 2.5 mm slice. The selection of planning CT slice thickness should therefore be considered if clinics decide to use automated registration functions.

CONCLUSION

The head and pelvis of a RANDO anthropomorphic phantom were used to find an optimal 1.0 ≤ Δτ (mm) ≤ 3.0 for the automatic registration procedure when using 3.0 mm slice spacing with planning CT. This study shows that using Δτ smaller than the reference set offers no advantage. Image noise is also a function of slice thickness. An optimal slice thickness range for reducing image noise of 2.5 ≤ Δτ (mm) ≤ 5.0 was also found using a CatPhan phantom. The overlap between these two ranges is 2.5 ≤ Δτ (mm) ≤ 3.0, providing optimal automatic registration accuracy and visibility.

ACKNOWLEDGEMENTS

This study was conducted with the support of the Ontario Institute for Cancer Research through funding provided by the government of Ontario.

References

  1. Woodford S, Yartsev S and Van Dyk J. Optimization of megavoltage CT scan registration settings for thoracic cases on helical tomotherapy. Phys Med Biol 2007; 52(15):N185�93.   [Medline]
  2. Broggi S, Di Muzio N and Fiorino C. Prostate tracking through daily MVCT-guided tomotherapy: Reduced motion by careful emptying of the rectum. Radiother Oncol 2007; 84(S1):S176.  
  3. Koller F and Roth J. Determination of the effective dose for CT examinations and influence of the setup parameters. Fortschr R�ntgenstr 2007; 179(1):36�43.   [Medline]
  4. Mori I. Antialiasing backprojection for helical MDCT. Med Phys 2008; 35(3): 1065�77.   [Medline]
  5. Sykes JR, Brettle DS, Magee DR and Thwaites DI. Investigation of uncertainties in image registration of cone beam CT to CT on an image-guided radiotherapy system. Phys Med Biol 2009; 54(24):7263�83.   [Medline]
  6. Ding GX, Duggan DM and Coffey CW. Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy. Phys Med Biol 2007; 52(6):1595�615.   [Medline]
  7. Murphy MJ, Wei Z, Fatyga M, Williamson J, Anscher M, Wallace T and Weiss E. How does CT image noise affect 3D deformable image registration for image-guided radiotherapy planning? Med Phys 2008; 35(3):1145�53.   [Medline]
  8. Li B, Avinash GB and Hsieh J. Resolution and noise trade off analysis for volumetric CT. Med Phys 2007; 34(10):3732�8.   [Medline]
  9. Fox TH, Huntzinger C, Johnstone PAS, Ogunleye T and Elder ES. Performance evaluation of an automated image registration algorithm using an integrated kV imaging and guidance system. J Appl Clin Med Phys 2006; 7(1):97�104.   [Medline]
  10. Abramoff MD, Magelhaes PJ and Ram SJ. Image Processing with ImageJ. Biophotonics Int 2004; 11(7):36�42.  
  11. Seet KYT, Barghi A, Yartsev S and Van Dyk J. The effects of field-of-view and patient size on CT numbers from cone-beam computed tomography. Phys Med Biol 2009; 54(20):1�12.   [Medline]

Received 22 october 2009; received in revised form 9 february 2010; accepted 23 february 2010

Correspondence: Department of Physics and Engineering, London Regional Cancer Program, London Health Sciences Centre, 790 Commissioners Road East, London, Ontario, Canada N6A 4L6. Tel.: Tel.: +1-5196858605; Fax: Fax: +1-5196858658; E-mail: slav.yartsev@lhsc.on.ca (Slav Yartsev).

Please cite as: Seet KYT, Barghi A, Yartsev S, Van Dyk J, Optimal slice thickness for cone-beam CT with on-board imager, Biomed Imaging Interv J 2010; 6(3):e31
<URL: http://www.biij.org/2010/3/e31/>


University of Malaya, Kuala Lumpur, Malaysia

Sponsors

GE
Bayer Healthcare
Siemens
Agfa
Toshiba
Philips
Elekta Fujifilm Barco Transmedic

Official publication of

ASEAN Association of Radiologists
ASEAN Society of Interventional Radiology
Asia-Oceania Federation of Organizations for Medical Physics
Asian Oceania Society of Radiology
College of Radiology, Academy of Medicine Malaysia
Southeast Asian Federation of Organisations of Medical Physics
South East Asian Association of Academic Radiologists

Published by

Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia




   

Biomedical Imaging and Intervention Journal. ISSN 1823-5530 RSS | Facebook | Twitter


 
   
Creative Commons License
Except where otherwise noted, articles published in the Biomedical Imaging and Intervention Journal
are distributed under the terms of the Creative Commons Attribution License, which permits
unrestricted use, distribution, and reproduction in any medium, provided the original work is properly
cited, including full bibliographic details and the URL, and this statement is included.