Hybrid imaging is the future of molecular imaging
RJ Hicks*, MD, FRACP,
EWF Lau, MBBS, FRANZCR,
DS Binns, ANMT
Centre for Molecular Imaging and Translational Medicine,
Peter MacCallum Cancer Centre, Victoria, Australia
Abstract
Correlative imaging has long been used in clinical
practice and particularly for the interpretation of nuclear medicine studies
wherein detailed anatomical information is often lacking. Previously,
side-by-side comparison or software co-registration techniques were applied but
suffered from technical limitations related to the differing geometries of the
imaging equipment, differences in the positioning of patients and displacement
of mobile structures between studies. The development of the first hybrid PET
and CT device struck a chord with the medical imaging community that is still
ringing loudly throughout the world. So successful has been the concept of
PET-CT that none of the major medical imaging manufacturers now offers
stand-alone PET scanners. Following close behind this success, SPECT-CT devices
have recently been adopted by the nuclear medicine community, already compelled
by the benefits of hybrid imaging through their experience with PET-CT. Recent
reports of adaptation of PET detectors to operate within the strong magnetic
field of MRI scanners have generated further enthusiasm. Prototype PET-MRI
devices are now in development. The complementary anatomical, functional and
molecular information provided by these techniques can now be presented in an
intuitive and aesthetically-pleasing format. This has made end-users more
comfortable with the results of functional imaging techniques than when the
same information is presented independently. Despite the primacy of anatomical
imaging for locoregional disease definition, the molecular characterisation
available from PET and SPECT offers unique complementary information for cancer
evaluation. A new era of cancer imaging, when hybrid imaging will be the
primary diagnostic tool, is approaching. © 2007 Biomedical Imaging and
Intervention Journal. All rights reserved.
Keywords: FDG, PET, SPECT, tomography, hybrid imaging,
oncology
Introduction
At medical schools throughout the world, doctors in
training are taught the basic sciences of anatomy, biology, physiology,
biochemistry, and so forth. However, when it comes to diagnostic imaging, there
is a heavy reliance placed on tutelage regarding anatomically-based techniques
while molecular and functional techniques are often given scant attention. Most
clinicians are therefore very comfortable with using CT, X-rays, ultrasound and
MRI because the output is readily understood, particularly if the images are
accompanied by judiciously placed arrows demonstrating the abnormalities.
Nuclear medicine techniques, apart from bone scanning, which resembles
sufficiently the mind’s eye perception of a skeleton to pass as an anatomical
representation, have often failed to capture the imagination of the clinician.
There is a feeling that functional imaging is somehow less valid than
radiological techniques because of the lack of fine spatial resolution. The
term “unclear medicine” is a common derogator used by clinicians when
discussing the work of nuclear medicine imaging specialists. Perhaps this
partly reflects the complexity of the principles that underpin these techniques
and that often do not lend themselves to “gestalt” interpretation.
Time-activity curves, deconvolutional analysis and compartmental modelling are
all based on mathematics and are generally held to be beyond common
understanding.
Nevertheless, the general public deals with functional
information on a daily basis and can integrate it into decision-making
processes if it is presented in an intuitive format. An obvious example appears
in weather reports on television or computer screens where serial radar images
of rainfall are displayed over a map of a geographical region with appropriate
labelling of landmarks. Different colours are used to display the intensity of
precipitation and a cinematic function is utilised to demonstrate the movement of a
storm system across the land. The end-user can easily decide whether to take an
umbrella on their walk, or to bring in their drying clothes from the washing-line
on the basis of such information without understanding the complexities of
radar, global positioning systems, satellites or meteorology. Hybrid imaging
provides a similar intuitive integration of information from functional and
structural imaging techniques without requiring a detailed understanding of the
technologies needed to produce them.
The history of hybrid imaging
Nuclear medicine actually started as a non-imaging
specialty with probes used to measure radioactivity in restricted regions of
the body, such as the thyroid gland. The first practical form of nuclear
medicine imaging was that performed on rectilinear scanners [1]. These devices
had a very small scintillation detector that moved methodically over the body
to produce, usually with a 1:1 scaling, dots on paper or, later, film
proportional to the number of radioactive events recorded in a particular unit
of time. Because the resulting images were “life-size”, they could be easily
overlaid on plain X-rays, which were also acquired without magnification. This
fusion of X-rays and rectilinear scans was used, for example, to diagnose
subphrenic collections prior to the availability of CT scanning (Figure 1).
Clearly, “anatamometabolic” imaging is not a new concept.
The development of the gamma camera by Hal Anger increased
the capacity for imaging larger regions of the body simultaneously but limited
the ability to perform direct correlations with anatomical images because the
film or computer-generated images were usually minified. However, with the
development and increasing clinical application of tomographic imaging
techniques such as CT and SPECT in the 1980s and 1990s, interest in image
fusion increased. Being computer-generated and therefore of an intrinsic
digital format, the images from these tomographic techniques were amenable to
software manipulations including translation, rotation, rescaling and,
potentially, to deformation. Because of the lack of deformation in brain
structures, software solutions for image co-registration were most successful
for neurological applications [2, 3]. However, even early in the development
of SPECT, fusion of anatomical and functional information was performed to
assess the distribution of radioligands that provided limited anatomical
information in their own right. Examples include radiolabeled monoclonal
antibodies [4, 5].
Relatively early in the development of clinical PET in
oncology the advantages of correlating structural and metabolic information
were also recognised [6]. PET co-registration with both CT and MRI found direct
clinical application in the guidance of neurosurgery [7]. It was also found to
improve the accuracy of staging lymph node involvement in patients with
non-small cell lung cancer (NSCLC) [8] and was used in radiotherapy planning
[9].
From the concept of image registration and fusion, it was
a relatively short intellectual leap of faith to develop hybrid scanners. GE
Medical Systems first commercially released a SPECT camera with an X-ray source
capable of providing low-dose CT images. This device, which went by the trade
name “Hawkeye”, immediately found application in oncology [10, 11]. Although
this iteration of SPECT-CT was available slightly earlier before PET-CT in a
commercial sense, the CT on the Hawkeye was of suboptimal diagnostic quality
and added significantly to the total acquisition time of a routine SPECT study.
The development of combined PET-CT systems integrating a diagnostic CT by
Townsend and co-workers has revolutionised hybrid imaging [12]. Early versions
of these hybrid devices incorporated partial-ring PET scanners but further
development has led to state-of-the-art PET and multi-detector CT (MDCT) now
being integrated into current devices [13]. The first combined PET-CT to be
available commercially was the Discovery LS scanner (GE Medical Systems, Milwaukee, WI). This device pragmatically involved simply bolting what were then current
generation PET and MDCT devices together and using a newly-designed patient bed
(Figure 2).
Recent Developments in Hybrid Imaging
Subsequent developments have seen the advent of PET-CT
designed as an integrated hardware and software platform. Such has been the
success of hybrid PET-CT scanners that none of the major manufacturers
currently offers stand-alone PET scanners for commercial sale. Recent advances
include incorporating 64-slice MDCT, new detector technologies and development
of time-of-flight scanners [14, 15]. The rapid take-up of PET-CT in clinical
practice has generated significant economies of scale to be realised in the
manufacture of systems and led to a significant fall in the cost of these
systems or availability of more technologically-advanced systems at comparable
prices. It has also stimulated reinvestment in instrumentation research and
development, which had stalled for many years as PET languished as a primarily
research tool with a low installed base. Accordingly, significant improvements
in performance with respect to resolution and scan acquisition times are
expected in the near future.
Following the success of PET-CT scanners with integrated
diagnostic MDCT, Philips and Siemens (Figure 3) both subsequently introduced
SPECT-CT scanners that also included a diagnostic MDCT. The superior diagnostic
quality of the CT on these scanners, particularly for soft tissues in the
abdomen, and the much shorter CT acquisition times offer significant advantages
over the GE Hawkeye system. However, these systems are more expensive and
involve higher radiation doses [16]. Hence the choice of system is highly dependent
on the range of applications for which the scanner will be used.
Despite the success and popularity of PET-CT and, more
recently, of SPECT-CT, there are some shortcomings in the use of CT as a
complementary anatomical imaging modality. Firstly, CT adds radiation dose to
the overall examination, particularly if used in a full diagnostic role [17].
Secondly, CT provides relatively poor soft tissue contrast in the absence of
oral and intravenous iodinated contrast, particularly if low-dose acquisition
protocols are utilised to minimise incremental radiation exposure. These two
theoretical limitations do not apply to MRI, which does not involve ionising
radiation and provides soft tissue imaging with high spatial resolution and
superior contrast compared to CT. MRI can also provide more advanced
‘functional’ techniques such as diffusion and perfusion imaging as well as
spectroscopy, which may be complementary to functional information obtained by
PET. Furthermore, the high sensitivity of PET may also complement the poor
signal strength inherent in current functional MRI imaging. The combination of
PET and MRI into a single scanner may therefore prove to be the ultimate hybrid
imaging modality, combining the metabolic and molecular information of PET with
the excellent anatomical detail of MRI, while offering new potential
applications with respect to functional MRI techniques [18].
There have been several recent publications on the value
of fused or co-registered PET-MRI images in pre-clinical and clinical practice
(19-22). There are, however, a number of technical problems to be overcome
before a hybrid PET-MRI scanner can become a reality. Both MRI and PET have the
potential to affect each other’s performance in their current form. One of the
main problems is that the photomultiplier tube, a fundamental component of
current PET detectors, will not function in a ‘magnet’ as the high magnetic
field causes electrons to deviate from their original trajectory, resulting in
loss of gain. A small prototype PET-MRI scanner has been developed using long
optical fibres to transport light from the detector to photomultiplier tubes
situated in a low field region [23].
A small 5.4 cm diameter MRI-compatible PET scanner
operating within the bore of a 9.4T MRI spectroscopy system and a 4.7T small
bore animal MRI imaging system have been developed successfully. There is
however significant light loss in the long optical fibres, leading to poor
energy and time resolution. This design is likely to be impractical for a large
number of detectors since the large volume of optical fibres required would
probably significantly limit the axial field of view. A more promising
alternative is the use of avalanche photodiode (APD) technology, which is a
compact and reliable silicon-based device. APDs have successfully replaced
photomultiplier tubes in a high resolution PET system, and can function in high
magnetic fields of up to 9.4T without any performance degradation. A
performance test of an APD-based LSO PET detector in a 7 T animal research MRI
scanner yielded encouraging results [24]. There was only slight degradation of
the PET detector energy spectra caused by magnetic gradients or RF pulses.
Current efforts in the development of PET-MRI scanner have
focused on small animal systems, because the technical requirements for such
systems are less demanding and applications are readily apparent [25]. However,
one commercial vendor, Siemens, has already showcased the first human PET-MRI
system at the Radiological Society of North America Meeting in 2006 and
preliminary imaging studies have been reported using a prototype PET-MRI[26].
The Siemens device uses a MR-compatible lutetium-oxalate (LSO) crystal-based
avalanche photodiode detector. The field of view in a combined mode with a PET
head insert is estimated to be about 20cm axially and 24cm radially. This means
that this system is primarily suitable for brain, and possibly other small
parts, imaging. There are many potentially useful PET-MRI applications in the
brain, including neurodegenerative disorders, epilepsy and tumours. Obviously,
at this time, the majority of data supporting such instrumentation comes from
cross-platform correlative studies. For example, a PET-MRI correlative study
reported positive relations between hippocampal atrophy and ipsilateral
association cortex hypometabolism in Alzheimer’s disease [27]. Digitally
performed PET-MRI coregistration was also found to increase childhood CNS
tumour characterisation in 90% of the cases and can be used to obtain a more
specific diagnosis with regard to tumour grading [20]. There was also a report
comparing the uptake of C11-choline PET tracer to the choline peak
on proton MR spectroscopy (MRS) in the assessment of brain tumours, suggesting
that they were both helpful in the differential diagnosis of cerebral lymphoma,
glioma and non-tumour lesions [28].
MRS imaging displays the relative concentrations of
chemical metabolites within a small volume of interest or voxel. In vivo
proton spectroscopy is most widely available and is used to look at biochemical
alterations in cancers and in characterisation of the ‘chemistry’ of target
lesions. It provides biochemical information that may be complementary to the
metabolic information obtained from PET, but unlike PET, does not expose patients
to ionising radiation. In clinical oncology, MRS was initially developed for
assessment of human brain tumours but has since been extended to evaluation of
prostate and breast cancers. The roles of proton MRS in oncology include
refinement of preoperative differential diagnosis data, which can be used to
guide surgical biopsy procedures, and detection and monitoring of treatment
response [29].
Traditionally, MRI scanning has been limited to assessment
of limited body regions due to prolonged imaging time and limited availability.
The new development of Total Imaging Matrix (TIM) technology allows fast,
high-resolution whole-body MRI imaging without the need for patient or surface
coil repositioning. The TIM used in the Siemens MAGNETOM AVANTO scanner combines
32 independent receiver channels with 76 array coil elements that can be
connected simultaneously [30]. With advancing technology and increasing demand
for high quality fast imaging, whole body human PET-MRI scanner might be
available in clinical practice in the not too distant future.
Hybrid imaging: Evolution or revolution?
It is easy to contend that the development of hybrid
imaging was merely an evolution of principles of correlative imaging developed
over the past 50 years. However, there is no doubt that this technology has
revolutionised the way that clinicians think about imaging. The fused image has
become the preferred visualisation tool of the end-user of imaging
investigations, the clinician. Although clearly better than contrast agents, the
superimposition of radiotracer signals on a set of CT or MRI images with which
the clinician feels comfortable, has increased confidence in the veracity of
the molecular information and its pathological or physiological basis.
In the case of PET-CT, besides the aesthetic advantages,
it has become clear that accurately co-registered images increase diagnostic
accuracy compared to independent or side-by-side reading of PET and CT data,
but more substantially increase the confidence with which abnormalities are
localised. The literature evaluating PET-CT is rapidly expanding but the vast
majority of preliminary and recent studies have attested to incremental
diagnostic value compared to PET or CT in a wide range of malignancies (31-44).
Better localisation of abnormalities can have significant management
implications, particularly in areas of complex anatomy [45]. Using the CT to
provide an attenuation map with which to correct emission data has been another
major advantage of hybrid PET-CT scanning. Previously various radioactive
sources including Ge-68 and Cs-137 [46] were used to determine the loss of
energy of detected annihilation photons due to tissue attenuation as they
passed through the body. Since PET image reconstruction relies currently on detection
of coincident detection of 2 photons, many photon pairs must pass through the
entire diameter of the body, creating significant degradation in sensitivity
for deep structures and restricting quantification of regional radioactivity.
Acquisition of transmission scans using radioactive sources was often a
time-consuming process, sometimes occupying as much as one third of the total
acquisition time in order to achieve adequate statistical quality to provide an
accurate map of tissue attenuation. Although innovative techniques including
simultaneous transmission and emission scanning [47] reduced this impost, the
ability to acquire a whole-body attenuation map in less than 1 minute using a
MDCT has dramatically reduced this component of a PET scan. Despite the higher
instrumentation costs, the more rapid scan acquisition protocols available on
current hybrid PET-CT scanners allows significantly higher throughput, allowing
economies of scale to maintain or reduce the unit cost of individual scans.
When the authors’ PET facility began operation in 1996, a
PET scan extending from the base of the brain to the mid-thigh required an
emission scan of close to 1 hour and a transmission scan of around 20 minutes
in duration. Since most studies were processed with both attenuation correction
and iterative reconstruction, the time taken to produce an image set for
analysis was often an hour or more. With such time constraints, the procedure
was limited to 6-8 patients per day. Today, the same axial extent of the body can
be scanned in less than 30 minutes. Some scanners are capable of less than
10-minute whole-body scan acquisition times. As a consequence, we are able to
perform 15 or more scans per scanner per day. This leads to more efficient use
of radiotracers with rapid radioactive decay, greater amortisation of equipment
and other fixed costs, such as maintenance. It also allows more productive use
of technologist, nursing, secretarial and medical staff.
Modern computing platforms also allow almost real-time
reconstruction of images. The advantages of this to patient comfort and
convenience are obvious. This has been reflected in a lower likelihood of
patient movement during the scan and ability to review scans while the patient
is still on the bed, so that additional images can be acquired if required.
This facility is routinely used in evaluating carcinoma of the stomach. A
whole-body scan is first acquired with the stomach empty. This allows
separation of the gastric wall from adjacent structures, including peri-gastric
lymph nodes. If there is no evidence of systemic metastasis on this scan, the
patient is administered buscopan as a smooth muscle relaxant and given 500 ml
of water to drink in order to distend the stomach. This leads to stretching of
the gastric smooth muscle, attenuating signal from it through partial volume
effects, while more clearly demarking the site of the primary lesion, which
being less compliant than the normal stomach, maintains its signal despite
distension (Figure 4).
The greater statistical quality of CT transmission maps
has also been advantageous in improving the quality of PET scans. However, the
fact that the attenuation characteristics of X-rays and annihilation photons
for varying tissues are significantly different means that correction factors
are required to translate from a CT-attenuation map to an appropriate 511 keV
map. This can lead to some discordance in quantitative analysis of tissue
tracer activity [48, 49]. Soft tissue attenuation is less of a problem for
SPECT reconstruction, since acquisition data in each projection are dominated
by photons arising closest to the detector and thereby passing through the
shortest distance of the body. Development of SPECT-CT systems has allowed more
robust attenuation correction to be performed. This is most helpful in the
abdomen, where the density of abdominal organs significantly reduces measured
activity from deep abdominal structures, particularly in obese patients.
However, because attenuation correction was not a routine part of standard
nuclear medicine and SPECT was often seen as an adjunct to planar imaging due
to the incremental acquisition times required, SPECT-CT has not resulted in
increased throughput capacity. Accordingly, SPECT-CT is more expensive than
standard nuclear medicine procedures because the equipment is more expensive
than a conventional gamma camera, and to optimally leverage the diagnostic
advantages of this technology it needs to be used primarily as a tomographic
device rather than for high throughput planar scanning. The relatively low
usage of the expensive CT component of the scanner could be viewed as wasteful.
However, it is believed that flexible imaging protocols that maximise the
benefits of hybrid SPECT-CT can be developed. These include having dedicated
planar scanners for whole-body screening and to acquire spot views. These would
then feed into a SPECT-CT scanner for dedicated regional imaging studies to
better define diagnostic questions that were inadequately resolved by planar
imaging.
In a hybrid PET-MRI system for whole body imaging, the
issue of attenuation correction for PET using MRI images will need to be
resolved. Segmentation and remapping algorithms will likely be required.
Disadvantages of hybrid technology
Although the CT and PET or SPECT components of hybrid
imaging studies are acquired contemporaneously, they are not acquired
simultaneously. Accordingly, this allows for movement to occur between the two
scans. The most common form of movement is that associated with normal respiration
[50]. It was recognised early in the development of PET that cardiac and
respiratory movement significantly degraded image quality and gating was
identified as a solution to this problem [51]. Because of the rapid acquisition
of CT images using a MDCT it is possible to acquire images during a single
breath-hold or even during normal breathing, fixing the position of structures
like the diaphragm, lungs, liver and spleen that move with respiration. The
emission images take minutes to acquire at each bed position and therefore
respiratory blurring of the image can occur. During normal tidal volume
breathing the lungs spend more of their time closer to an end-expiratory volume
than to an end-inspiratory volume and certainly not at the degree of expansion
associated with a forced inspiration, which is the preferred state for
diagnostic CT of the chest. Thus, comparison of the CT images with the
integrated PET emission image demonstrates a degree of misregistration at the
level of the diaphragm although this infrequently causes major diagnostic
problems [52]. The most common finding is that CT structures appear to be lower
than the corresponding metabolic signal. This misregistration is also
manifested in the assignment of the attenuation characteristics to structures
at around the level of the diaphragm. By incorrectly assuming counts that have
actually arisen within dense liver parenchyma came from aerated lung based on
the CT-attenuation map, these counts are given less weighting and therefore
appear as an area of relative photopenia on reconstructed images (Figure 5).
Various manoeuvres have been tried to eliminate these misregistration artefacts.
These have included altering breath-holding to mid-inspiration or
end-expiration during acquisition of the CT component. It has been found that
any instruction regarding breathing tends to significantly alter breathing
pattern and therefore the authors have chosen to simply acquire the CT without
any instructions to the patient other than to lie still. For situations where
extremely accurate registration of anatomical and structural information is
required, respiratory gating of both the CT and PET components may be an option
but it places time constraints on the study and requires a more sophisticated
set-up. Nevertheless, this may be worthwhile, particularly for planning
treatment of basal lung cancers. Techniques have been developed to allow what
is termed 4d PET-CT [53].
Gross physical movements can also occur. Making the
patient aware of the need to remain still, efforts to make them as comfortable
as possible and to reduce total scan time are, in the authors’ experience, the
most efficacious methods of reducing patient movement. Some facilities,
however, use body restraints that minimise the capacity for patient movement.
Even with physical restraints, patient movements can occur and give rise to
both attenuation artefacts and diagnostic difficulties related to
misregistration of PET and CT data.
Unlike PET-CT, where PET and CT imaging are acquired
sequentially, simultaneous acquisition of PET data and MRI images is possible
in a hybrid PET-MRI scanner as the PET scanner operates within the bore of the
MRI magnet. This provides for the first time simultaneous anatomical and
functional imaging with not only potentially perfect fusion PET-MRI images but
also the prospect of performing dynamic imaging to obtain valuable functional
information.
Should hybrid imaging be an adjunct to or replacement of anatomical
imaging?
Because of the relatively high cost of PET and its restricted
availability in many parts of the world, PET has been generally reserved for
cases with equivocal conventional imaging results. However, in the authors’
experience, the major impact of FDG PET is to prevent futile procedures in
patients in whom PET detects metastatic disease unrecognised by conventional
staging techniques irrespective of whether there was any equivocation on these
tests. Since most oncological therapies are selected and monitored based on the
presence and extent of apparent disease, more accurate definition of these
parameters is important to appropriate treatment delivery. For example, in
radiotherapy patients, for whom adequate coverage of gross tumour volume is
vital to achieving local control and survival, better definition of macroscopic
regional nodal disease and exclusion of distant metastases would be expected to
improve patient outcomes [54]. More precise characterisation of the nature and
location of focal FDG accumulations identified by PET-CT is likely to further
improve diagnostic performance and thereby, treatment selection and planning.
By avoiding unnecessary or futile surgery, radiotherapy or chemotherapy and
better assessing the response to these treatments, there is substantial
opportunity to not only improve patient care but also to reduce costs, despite
the higher upfront cost of the imaging component of the management paradigm if
PET-CT and other hybrid imaging tests were used as the primary diagnostic test.
There is increasing evidence across a broad range of indications that hybrid
PET-CT [55] and hybrid SPECT-CT (56-61) are usually more accurate that either
modality allow and not infrequently when compared to side-by-side comparison of
each modality acquired separately. The huge number of studies being published
using these technologies will continue to refine the clinical performance and
role of hybrid imaging but there is no doubt that there is no turning back. The
future of cancer imaging lies with hybrid imaging technologies. Because of the
extremely high contrast associated with many SPECT tracers (Figure 6), the
authors find that the addition of CT is even more valuable than it is to FDG
PET wherein there is already substantial vicarious anatomical information by
way of uptake in normal tissues. An emerging application of SPECT-CT will be
the development of algorithms for radionuclide therapy dosimetry planning
(Figure 7).
Post-script note: Although no mention of optical
imaging is made above, there is preliminary work on tomographic imaging devices
based on this technology [62]. The primary author is of the opinion that
optical imaging will have increasing clinical application and may one day also
be the subject of clinical hybrid imaging platforms.
References
-
Blahd WH. Ben Cassen and the development of the rectilinear scanner. Semin Nucl Med 1996; 26(3):165-70.
[Medline]
-
Rizzo G, Gilardi MC, Prinster A et al. A bioimaging integration system implemented for neurological applications. J Nucl Biol Med 1994; 38(4):579-85.
[Medline]
-
Pietrzyk U, Herholz K, Schuster A et al. Clinical applications of registration and fusion of multimodality brain images from PET, SPECT, CT, and MRI. Eur J Radiol 1996; 21(3):174-82.
[Medline]
-
Kramer EL, Noz ME, Sanger JJ et al. CT-SPECT fusion to correlate radiolabeled monoclonal antibody uptake with abdominal CT findings. Radiology 1989; 172(3):861-5.
[Medline]
-
Koral KF, Zasadny KR, Kessler ML et al. CT-SPECT fusion plus conjugate views for determining dosimetry in iodine-131-monoclonal antibody therapy of lymphoma patients. J Nucl Med 1994; 35(10):1714-20.
[Medline]
-
Wahl RL, Quint LE, Cieslak RD et al. "Anatometabolic" tumor imaging: fusion of FDG PET with CT or MRI to localize foci of increased activity. J Nucl Med 1993; 34(7):1190-7.
[Medline]
-
Kraus GE, Bernstein TW, Satter M et al. A technique utilizing positron emission tomography and magnetic resonance/computed tomography image fusion to aid in surgical navigation and tumor volume determination. J Image Guid Surg 1995; 1(6):300-7.
[Medline]
-
Vansteenkiste JF, Stroobants SG, Dupont PJ et al. FDG-PET scan in potentially operable non-small cell lung cancer: do anatometabolic PET-CT fusion images improve the localisation of regional lymph node metastases? The Leuven Lung Cancer Group. Eur J Nucl Med 1998; 25(11):1495-501.
[Medline]
-
Cai J, Chu JC, Recine D et al. CT and PET lung image registration and fusion in radiotherapy treatment planning using the chamfer-matching method. Int J Radiat Oncol Biol Phys 1999; 43(4):883-91.
[Medline]
-
Delbeke D, Sandler MP. The role of hybrid cameras in oncology. Semin Nucl Med 2000; 30(4):268-80.
[Medline]
-
Israel O, Keidar Z, Iosilevsky G et al. The fusion of anatomic and physiologic imaging in the management of patients with cancer. Semin Nucl Med 2001; 31(3):191-205.
[Medline]
-
Beyer T, Townsend DW, Brun T et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000; 41(8):1369-79.
[Medline]
-
Beyer T, Townsend DW. Putting 'clear' into nuclear medicine: a decade of PET/CT development. Eur J Nucl Med Mol Imaging 2006; 33(8):857-61.
[Medline]
[CrossRef]
-
Muehllehner G, Karp JS. Positron emission tomography. Phys Med Biol 2006; 51(13):R117-37.
[Medline]
[CrossRef]
-
Surti S, Kuhn A, Werner ME et al. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007; 48(3):471-80.
[Medline]
-
O'Connor MK, Kemp BJ. Single-photon emission computed tomography/computed tomography: basic instrumentation and innovations. Semin Nucl Med 2006; 36(4):258-66.
[Medline]
[CrossRef]
-
Wu TH, Huang YH, Lee JJ et al. Radiation exposure during transmission measurements: comparison between CT- and germanium-based techniques with a current PET scanner. Eur J Nucl Med Mol Imaging 2004; 31(1):38-43.
[Medline]
[CrossRef]
-
Seemann MD. Whole-body PET/MRI: the future in oncological imaging. Technol Cancer Res Treat 2005; 4(5):577-82.
[Medline]
-
Hayakawa N, Uemura K, Ishiwata K et al. A PET-MRI registration technique for PET studies of the rat brain. Nucl Med Biol 2000; 27(2):121-5.
[Medline]
-
Borgwardt L, Hojgaard L, Carstensen H et al. Increased fluorine-18 2-fluoro-2-deoxy-D-glucose (FDG) uptake in childhood CNS tumors is correlated with malignancy grade: a study with FDG positron emission tomography/magnetic resonance imaging coregistration and image fusion. J Clin Oncol 2005; 23(13):3030-7.
[Medline]
[CrossRef]
-
Seemann MD, Meisetschlaeger G, Gaa J et al. Assessment of the extent of metastases of gastrointestinal carcinoid tumors using whole-body PET, CT, MRI, PET/CT and PET/MRI. Eur J Med Res 2006; 11(2):58-65.
[Medline]
-
Ruf J, Lopez Hanninen E, Bohmig M et al. Impact of FDG-PET/MRI image fusion on the detection of pancreatic cancer. Pancreatology 2006; 6(6):512-9.
[Medline]
[CrossRef]
-
Marsden PK, Strul D, Keevil SF et al. Simultaneous PET and NMR. Br J Radiol 2002; 75 Spec No:S53-9.
[Medline]
-
Pichler BJ, Judenhofer MS, Catana C et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med 2006; 47(4):639-47.
[Medline]
-
Lucignani G. Time-of-flight PET and PET/MRI: recurrent dreams or actual realities? Eur J Nucl Med Mol Imaging 2006; 33(8):969-71.
[Medline]
[CrossRef]
-
Catana C, Wu Y, Judenhofer MS et al. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 2006; 47(12):1968-76.
[Medline]
-
Meguro K, LeMestric C, Landeau B et al. Relations between hypometabolism in the posterior association neocortex and hippocampal atrophy in Alzheimer's disease: a PET/MRI correlative study. J Neurol Neurosurg Psychiatry 2001; 71(3):315-21.
[Medline]
-
Utriainen M, Komu M, Vuorinen V et al. Evaluation of brain tumor metabolism with [11C]choline PET and 1H-MRS. J Neurooncol 2003; 62(3):329-38.
[Medline]
-
Kwock L, Smith JK, Castillo M et al. Clinical role of proton magnetic resonance spectroscopy in oncology: brain, breast, and prostate cancer. Lancet Oncol 2006; 7(10):859-68.
[Medline]
[CrossRef]
-
Gaa J, Rummeny EJ, Seemann MD. Whole-body imaging with PET/MRI. Eur J Med Res 2004; 9(6):309-12.
[Medline]
-
Hany TF, Steinert HC, Goerres GW et al. PET diagnostic accuracy: improvement with in-line PET-CT system: initial results. Radiology 2002; 225(2):575-81.
[Medline]
-
Antoch G, Stattaus J, Nemat AT et al. Non-small cell lung cancer: dual-modality PET/CT in preoperative staging. Radiology 2003; 229(2):526-33.
[Medline]
[CrossRef]
-
Bar-Shalom R, Yefremov N, Guralnik L et al. Clinical performance of PET/CT in evaluation of cancer: additional value for diagnostic imaging and patient management. J Nucl Med 2003; 44(8):1200-9.
[Medline]
-
Bristow RE, del Carmen MG, Pannu HK et al. Clinically occult recurrent ovarian cancer: patient selection for secondary cytoreductive surgery using combined PET/CT. Gynecol Oncol 2003; 90(3):519-28.
[Medline]
-
Cohade C, Osman M, Marshall LN et al. PET-CT: accuracy of PET and CT spatial registration of lung lesions. Eur J Nucl Med Mol Imaging 2003; 30(5):721-6.
[Medline]
[CrossRef]
-
Kantorova I, Lipska L, Belohlavek O et al. Routine (18)F-FDG PET preoperative staging of colorectal cancer: comparison with conventional staging and its impact on treatment decision making. J Nucl Med 2003; 44(11):1784-8.
[Medline]
-
Cerfolio RJ, Ojha B, Bryant AS et al. The accuracy of integrated PET-CT compared with dedicated PET alone for the staging of patients with nonsmall cell lung cancer. Ann Thorac Surg 2004; 78(3):1017-23; discussion 1017-23.
[Medline]
[CrossRef]
-
Even-Sapir E, Parag Y, Lerman H et al. Detection of recurrence in patients with rectal cancer: PET/CT after abdominoperineal or anterior resection. Radiology 2004; 232(3):815-22.
[Medline]
[CrossRef]
-
Grisaru D, Almog B, Levine C et al. The diagnostic accuracy of 18F-fluorodeoxyglucose PET/CT in patients with gynecological malignancies. Gynecol Oncol 2004; 94(3):680-4.
[Medline]
[CrossRef]
-
Wahl RL. Why nearly all PET of abdominal and pelvic cancers will be performed as PET/CT. J Nucl Med 2004; 45 Suppl 1:82S-95S.
[Medline]
-
Chen YK, Su CT, Ding HJ et al. Clinical usefulness of fused PET/CT compared with PET alone or CT alone in nasopharyngeal carcinoma patients. Anticancer Res 2006; 26(2B):1471-7.
[Medline]
-
De Wever W, Ceyssens S, Mortelmans L et al. Additional value of PET-CT in the staging of lung cancer: comparison with CT alone, PET alone and visual correlation of PET and CT. Eur Radiol 2007; 17(1):23-32.
[Medline]
[CrossRef]
-
Sebastian S, Lee SI, Horowitz NS et al. PET-CT vs. CT alone in ovarian cancer recurrence. Abdom Imaging 2007.
[Medline]
[CrossRef]
-
Shin SS, Jeong YY, Min JJ et al. Preoperative staging of colorectal cancer: CT vs. integrated FDG PET/CT. Abdom Imaging 2007.
[Medline]
[CrossRef]
-
Lau WF, Binns DS, Ware RE et al. Clinical experience with the first combined positron emission tomography/computed tomography scanner in Australia. Med J Aust 2005; 182(4):172-6.
[Medline]
-
Mirzaei S, Guerchaft M, Bonnier C et al. Use of segmented CT transmission map to avoid metal artifacts in PET images by a PET-CT device. BMC Nucl Med 2005; 5(1):3.
[Medline]
[CrossRef]
-
Meikle SR, Bailey DL, Hooper PK et al. Simultaneous emission and transmission measurements for attenuation correction in whole-body PET. J Nucl Med 1995; 36(9):1680-8.
[Medline]
-
Kamel E, Hany TF, Burger C et al. CT vs 68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube current. Eur J Nucl Med Mol Imaging 2002; 29(3):346-50.
[Medline]
-
Nakamoto Y, Osman M, Cohade C et al. PET/CT: comparison of quantitative tracer uptake between germanium and CT transmission attenuation-corrected images. J Nucl Med 2002; 43(9):1137-43.
[Medline]
-
Goerres GW, Burger C, Kamel E et al. Respiration-induced attenuation artifact at PET/CT: technical considerations. Radiology 2003; 226(3):906-10.
[Medline]
[CrossRef]
-
Ter-Pogossian MM, Bergmann SR, Sobel BE. Influence of cardiac and respiratory motion on tomographic reconstructions of the heart: implications for quantitative nuclear cardiology. J Comput Assist Tomogr 1982; 6(6):1148-55.
[Medline]
-
Osman MM, Cohade C, Nakamoto Y et al. Clinically significant inaccurate localization of lesions with PET/CT: frequency in 300 patients. J Nucl Med 2003; 44(2):240-3.
[Medline]
-
Nehmeh SA, Erdi YE, Pan T et al. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys 2004; 31(12):3179-86.
[Medline]
-
Hicks RJ, Mac Manus MP. 18F-FDG PET in candidates for radiation therapy: is it important and how do we validate its impact? J Nucl Med 2003; 44(1):30-2.
[Medline]
-
Czernin J, Allen-Auerbach M, Schelbert HR. Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. J Nucl Med 2007; 48 Suppl 1:78S-88S.
[Medline]
-
Schillaci O, Danieli R, Manni C et al. Is SPECT/CT with a hybrid camera useful to improve scintigraphic imaging interpretation? Nucl Med Commun 2004; 25(7):705-10.
[Medline]
-
Horger M, Eschmann SM, Pfannenberg C et al. Evaluation of combined transmission and emission tomography for classification of skeletal lesions. AJR Am J Roentgenol 2004; 183(3):655-61.
[Medline]
-
Hillel PG, van Beek EJ, Taylor C et al. The clinical impact of a combined gamma camera/CT imaging system on somatostatin receptor imaging of neuroendocrine tumours. Clin Radiol 2006; 61(7):579-87.
[Medline]
[CrossRef]
-
Roach PJ, Schembri GP, Ho Shon IA et al. SPECT/CT imaging using a spiral CT scanner for anatomical localization: Impact on diagnostic accuracy and reporter confidence in clinical practice. Nucl Med Commun 2006; 27(12):977-87.
[Medline]
[CrossRef]
-
Sodee DB, Sodee AE, Bakale G. Synergistic value of single-photon emission computed tomography/computed tomography fusion to radioimmunoscintigraphic imaging of prostate cancer. Semin Nucl Med 2007; 37(1):17-28.
[Medline]
[CrossRef]
-
Schillaci O, Danieli R, Filippi L et al. Scintimammography with a hybrid SPECT/CT imaging system. Anticancer Res 2007; 27(1B):557-62.
[Medline]
-
Hielscher AH, Bluestone AY, Abdoulaev GS et al. Near-infrared diffuse optical tomography. Dis Markers 2002; 18(5-6):313-37.
[Medline]
Received 25 September 2007; accepted 6 October 2007
Correspondence: Centre for Molecular Imaging and Translational Medicine, Peter MacCallum Cancer Centre, 12 Cathedral Place, East Melbourne, Victoria, Australia. Tel.: +61-3-9656-1852; Fax: +61-3-9656-1826; E-mail: rod.hicks@petermac.org (Rodney Hicks).
Please cite as: Hicks RJ, Lau EWF, Binns DS,
Hybrid imaging is the future of molecular imaging, Biomed Imaging Interv J 2007; 3(3):e49